HCMU Specifications
HCMU measures: The efficacy of any pump or compressor (Using current coils - non-invasive). |
|
Pump/Compressor Measurements Specifications (Current Coil Per phase) -Three Channels: |
|
Measurement Range |
0-80 A Max |
Measurement Resolution |
.1 A |
Measurement Accuracy |
2% Max |
Threshold for detection |
>1.5 A for All phases. |
Alarm LED |
ON if pump or compressor is off |
Current Coils Specifications |
The same specifications like CMB current coils |
Interface & Powering of HCMU Specifications: |
|
Interface to Main Board |
RS-485 |
Power |
12 volt – Internally form Main Board |
Indicators |
Two LEDS (online ,Alarm) |
Environmental & Mechanical Specifications: |
|
Enclosure |
ABS Plastic |
Operating Temperature |
0-50 °C |
Storage Temperature |
0-70 °C |
Relative Humidity |
0-95% RH |
Weight |
320 gram |